Canada Gazette, Part I, Volume 150, Number 25: Regulations Amending the Radiation Emitting Devices Regulations (Dental X-ray Equipment)

June 18, 2016

Statutory authority

Radiation Emitting Devices Act

Sponsoring department

Department of Health

REGULATORY IMPACT ANALYSIS STATEMENT

(This statement is not part of the Regulations.)

Issues

Dental radiography is one of the most valuable tools in the field of dental care, supporting the diagnosis, treatment and management of dental health concerns. It is estimated that nearly 14 million Canadians receive dental X-rays each year, a process that involves exposure to ionizing radiation. Since this type of radiation is capable of damaging the deoxyribonucleic acid (DNA) of healthy cells, which may potentially lead to cancer, the radiation safety of dental X-ray equipment imported, sold and leased in Canada must be carefully managed.

The widely adopted model for protection from ionizing radiation risk assumes that the probability of cancer from radiation exposure is proportional to the dose received. Consequently, the risk of harm from dental X-rays is generally very low for individuals — given that the radiation dose is low for each exposure — relative to other types of medical imaging examinations. From a public health standpoint, however, the risk is more significant due to the large number of Canadians receiving dental X-rays. To help protect the Canadian population as a whole, dental X-ray equipment imported, sold and leased in Canada must meet requirements of the Radiation Emitting Devices Act (REDA) and the Radiation Emitting Devices Regulations.

The Radiation Emitting Devices Regulations (the Regulations) are no longer aligned with the current international standards of the International Electrotechnical Commission (IEC), updated in 2012. Amending the Regulations to align with these standards is necessary to strengthen Canada's regulatory system and

Strengthen radiation safety requirements

Ionizing radiation, such as X-rays, can result in deleterious health effects. Canada's regulations will be more effective in protecting operators and patients of dental X-ray equipment if they reflect the latest scientific information related to radiation safety requirements. Current Regulations are no longer aligned with the current IEC Standards which are a benchmark for industry. This represents a gap within Canada's current regulatory framework and could put operators and patients at risk of receiving unnecessary doses of radiation during the operation of dental X-ray equipment.

Address a broader scope of dental X-ray technologies

Advancements in dental X-ray technologies have resulted in additional types of dental X-ray equipment coming into Canada (e.g. CBCT). The Regulations do not specifically address the current breadth of technology available to dental professionals. Without proper regulatory oversight of new technologies and technological advancements in existing technologies, operators and patients could be exposed unnecessarily to doses of radiation that are higher than needed.

In addition, transportable hand-held dental X-ray equipment is not available for use in Canada as it is currently not compliant with the radiation safety requirements of the Regulations. However, transportable hand-held dental X-ray equipment would provide dental practitioners access to another diagnostic tool, for use in situations where traditional wall- or stand-mounted dental X-ray equipment is not feasible (i.e. emergency medicine). The proposed regulatory amendments would address new technologies such as dental volumetric reconstruction technologies (e.g. CBCT) and transportable hand-held X-ray equipment.

Provide more information to support optimized equipment use

Optimization involves using the lowest dose of radiation necessary for a given clinical purpose (e.g. to produce an X-ray image that is of adequate diagnostic quality). Current Regulations require information necessary for the safe and proper operation of the equipment but do not have specific requirements for information on optimization. To address this issue, the proposed amendments would require the manufacturer, importer, or distributor of dental X-ray equipment to provide additional information on equipment design, available settings and their impact on radiation output, instructions on their use and quality control. This will provide operators with the means to optimize equipment use, thereby protecting both themselves and patients from unnecessary radiation exposure.

Background

In Canada, radiation protection in dentistry is a shared responsibility between the federal, provincial and territorial governments as well as the dental professionals involved in the delivery of dental care. Health Canada administers legislation applicable to the importation and sale of dental X-ray equipment under the REDA and the Regulations, as well as the Food and Drugs Act and the Medical Devices Regulations. Provincial and territorial governments and professional dental associations have made rules pertaining to the installation and safe use of this equipment.

The Consumer and Clinical Radiation Protection Bureau (CCRPB) within Health Canada administers the REDA and its Regulations in order to help protect the health of Canadians from potential hazards associated with radiation emitting devices, including dental X-ray equipment. The REDA includes provisions respecting the sale (including re-sale), lease and importation of radiation emitting devices, and authorizes the making of regulations pertaining to, among other things, the labelling, packaging, advertising, construction and functionality of radiation emitting devices.

Schedule II, Part II (Dental X-ray Equipment with an Extra-oral Source) of the Regulations prescribes particular radiation safety standards for dental X-ray equipment, and was last amended on May 5, 1993.

Currently, the Regulations provide general regulatory oversight to wall-mounted, intra-oral equipment, cephalometric and panoramic dental X-ray equipment but they do not adequately address the technological advancements that have occurred with this equipment. They permit the importation of CBCT dental X-ray equipment but prohibit the importation and sale of hand-held dental X-ray equipment. The latter does not meet current safety requirements in the Regulations that require the irradiation switch to allow the operator to stand at least 3 m from the device.

More sophisticated types of wall-mounted, cephalometric and panoramic dental X-ray equipment are coming to Canada and although CBCT equipment is permitted under the current Regulations, there is nothing specifically that addresses the innovations respecting this technology within the Regulations. Aligning the Regulations with aspects of the 2012 IEC Standards would enhance Canada's radiation safety requirements for these types of equipment, thereby providing greater protection to patients and operators. The proposed Regulations would also permit the importation and sale of transportable hand-held dental X-ray equipment, with provisions for safety requirements specific to hand-held devices.

Objectives

The objectives of the proposed regulatory amendments are to strengthen radiation safety requirements, in alignment with IEC Standards, for new equipment (i.e. equipment manufactured after the coming-into-force date of the amended Regulations), modern equipment (i.e. more sophisticated types of equipment currently permitted, but not adequately addressed under the existing Regulations), and new technology (i.e. transportable hand-held equipment that is currently prohibited under the existing Regulations); broaden the scope of the Regulations to address new and modern dental X-ray technologies; and require the provision of information to support optimization of equipment use.

There is a need to update the regulatory requirements for dental X-ray equipment in order to enable Health Canada to exercise appropriate regulatory oversight on the sale, resale, lease, and importation of dental X-ray equipment in Canada to ultimately enhance the protection of operators and patients from the radiation emitted by those devices.

Description

The Regulations set out requirements respecting information, labelling, construction, and functioning of radiation emitting devices.

The proposed amendments would update such requirements for dental X-ray equipment, including conventional wall-mounted and mobile dental X-ray equipment, panoramic equipment, cephalometric equipment, CBCT equipment and transportable hand-held dental X-ray equipment, as indicated below.

Update information and labelling requirements for dental X-ray equipment

The proposed amendments would require manufacturers to provide additional information for each piece of dental X-ray equipment, including quality control procedures and dosimetric information. These information requirements are more extensive than what is currently addressed in the Regulations and provide operators with the ability to properly operate and maintain their machines, ensuring consistent and appropriate patient dose delivery.

The proposed amendments would furthermore introduce a new requirement for a label on hand-held dental X-ray equipment. The label would inform operators of radiation safety considerations associated with hand-held equipment and be intended to promote safe use and mitigate radiation risks to operators.

Update construction standards

Requirements in the Regulations would be amended with respect to such features as minimum tube voltage setting (increasing from 50 kV to 60 kV), exit field size for intra-oral devices (reducing the maximum diameter from 7 cm to 6 cm), radiation absorbing filters (increasing the minimum attenuation requirements for extra-oral devices), X-ray field limitations (e.g. requirements specific to CBCT machines) and focal spot to skin distance (increased to 20 cm for intra-oral devices). These changes would reduce the radiation risk to which the patient and operator are exposed during operation of the X-ray device.

In addition, the proposed amendments would enable the importation and sale of transportable hand-held intra-oral dental devices in Canada while introducing requirements to mitigate radiation risks from these devices.

Update functioning standards

Functioning standards are criteria that require equipment to perform within certain specified parameters under normal conditions of use. The proposed regulatory amendments would update functioning standards, including requirements for improved accuracy of loading factors, and a more restrictive limit on leakage radiation, which would help to protect operators and patients from receiving unnecessary doses of radiation from dental X-ray equipment.

“One-for-One” Rule

The new standards set out requirements pertaining to labelling, construction, functioning, and information to accompany the device for its safe and proper operation, but do not introduce any activities that are considered to be an administrative burden to importers and sellers. There is also limited manufacturing of these devices in Canada. Therefore, the “One-for-One” Rule does not apply to this proposal.

Small business lens

The small business lens does not apply because the nationwide cost impacts of these proposed amendments are less than $1 million annually. The majority of manufacturers of dental X-ray equipment have already aligned their construction and function specifications with international standards (i.e. IEC Standards). Therefore, any cost associated with the proposed amendments is expected to be minor.

Consultation

Health Canada conducted a targeted stakeholder consultation from August to November 2013 to solicit early input from stakeholders on potential amendments to the Regulations. Comments were received from a total of 13 separate stakeholders, including representation from provincial government radiation authorities, the dental industry, dental professionals associations, federal government departments, academia and a private radiation safety company.

Health Canada's approach to align its Regulations with international equipment standards (i.e. IEC Standards) received strong support from the stakeholders. Some stakeholders (including provincial government radiation authorities) sought guidance on CBCT, such as quality control requirements and training for operators. To address this, the proposed amendments would require manufacturers, distributors and importers to provide quality control information on all dental X-ray equipment for sale and lease in Canada, including dental volumetric reconstruction equipment.

Transportable hand-held dental X-ray equipment was identified by some stakeholders as a tool that could potentially provide additional benefits to Canadians in certain situations, with respect to patient accessibility, when compared to transportable mobile dental X-ray equipment that is currently authorized for sale and import in Canada. However, support was tempered by the stipulation that they only be used when the use of a fixed or mobile unit was not feasible (e.g. transportable hand-held devices should not be used in standard dental offices). The proposed amendments would permit the importation, sale and lease of transportable hand-held dental X-ray equipment in Canada and would set out requirements to mitigate risks from these devices such as the inclusion of a second irradiation switch that gives the operator the option to be at least 2 m from the X-ray source assembly when the equipment is supported on a stand.

Rationale

The proposed amendments would update Canada's regulatory oversight for radiation safety on new and modern dental X-ray equipment technologies and align Canada's regulations with the current IEC Standards, which are already employed by the United States and the European Union. This would help strengthen radiation safety for patients and operators by requiring that all dental X-ray equipment imported and sold in Canada support optimized radiation doses to patients and reduced exposures to operators. To achieve this, the proposed amendments would require manufacturers, importers and distributors to meet equipment requirements for information, labelling, construction and performance.

As the quality and construction of dental X-ray equipment may vary across manufacturers, some of these differences may put operators and patients at increased risk for higher received doses. Updating the regulatory requirements so that devices imported and sold in Canada meet internationally established standards for safety and performance would help ensure optimized radiation doses to patients and operators.

To illustrate how the proposed amendments help optimize radiation doses to patients and operators, the proposed amendments would limit the X-ray field for intra-oral devices from the current limit of 7 cm diameter to no more than 6 cm, thus reducing the dose area product by about 25%. The dose area product is a measure that takes into account both radiation dose and how much of the body is receiving the radiation, and therefore is relative to radiation risk. The proposed amendments would also require accompanying documents to include a method to determine the dose area product from the device, as well as quality control information, both of which would help the operator to ensure the settings on the dental X-ray device are producing an appropriate dose area product on an ongoing basis. Ultimately, this would help operators protect themselves and their patients from exposure to unnecessary doses of radiation.

The proposed amendments would also provide operators with the opportunity to have access to hand-held dental X-ray equipment which may offer benefits in specific situations where the use of traditional wall- or stand-mounted devices is not feasible (i.e. emergency medicine).

In addition to increasing radiation safety requirements, the proposed amendments would offer economic benefits to Canada. Due to discrepancies that exist between the radiation safety requirements of the current Canadian dental X-ray regulations and the latest IEC Standards, manufacturers and importers of certain types of dental X-ray equipment (e.g. transportable hand-held dental X-ray devices) are currently not importing or selling these devices in Canada as modifications would be required in order to comply with the Canadian regulations. As nearly all dental X-ray equipment is manufactured outside of Canada and is designed to meet IEC Standards, the required modifications to import or sell certain devices in Canada currently result in a technical barrier and create additional costs which may be passed on to the consumer. By aligning with the current IEC Standards, the proposed amendments will reduce the technical barrier to manufacturers, importers and distributors that do not currently sell or import products in Canada because they do not wish to modify their products.

In terms of the potential costs to business (industry) as a result of the regulatory proposal, it is estimated that compliance costs for implementing the proposed amendments are minimal. Based on Innovation, Science and Economic Development Canada trade data from 2013, approximately 2 000 dental X-ray devices are imported into Canada per year. Given that the majority of dental X-ray devices are compliant with IEC Standards, it is estimated that no more than 25% of those devices would be non-compliant with the proposed Regulations. The costs associated with bringing a device into compliance, which include generation of new content, translations, and printing of manuals, are approximately $60; therefore, it is roughly estimated that the compliance costs for implementing the proposed amendments would be approximately CAN$30,000 per year. As most dental X-ray equipment is already manufactured to meet the current IEC Standards, the majority of the costs associated with this proposal are expected to relate to updating the information contained in manuals and labels, as well as some minor modifications to the construction of the devices.

Implementation, enforcement and service standards

In order to assist the dental industry transition, the proposed Regulations would come into force six months after the day that they are published in the Canada Gazette, Part II, allowing manufacturers sufficient time to align themselves with the amended Regulations.

The proposed Regulations would not alter existing compliance mechanisms. Compliance and enforcement would continue to be undertaken by Health Canada inspectors under the authority of the REDA and its regulations.

The current Regulations will continue to apply to dental X-ray equipment that was manufactured before the day on which these proposed amendments to the Regulations come into force.

Contact

Tara Bower
Director
Office of Science Policy, Liaison and Coordination
Environmental and Radiation Health Sciences Directorate
Healthy Environments and Consumer Safety Branch
Health Canada
Address Locator 4908D
269 Laurier Avenue West
Ottawa, Ontario
K1A 0K9
Fax: 613-952-5397
Email: ccrpb-pcrpcc@hc-sc.gc.ca

PROPOSED REGULATORY TEXT

Notice is given, pursuant to subsection 13(2) of the Radiation Emitting Devices Act (see footnote a), that the Governor in Council, pursuant to subsection 13(1) of that Act, proposes to make the annexed Regulations Amending the Radiation Emitting Devices Regulations (Dental X-ray Equipment).

Interested persons may make representations concerning the proposed Regulations within 75 days after the date of publication of this notice. All such representations must cite the Canada Gazette, Part I, and the date of publication of this notice, and be addressed to Tara Bower, Director, Office of Science Policy, Liaison and Coordination, Department of Health, Postal Locator: 4908D, 269 Laurier Avenue W, Room 8-016, Ottawa, Ontario K1A 0K9 (fax: 613-952-5397; email: ccrpb-pcrpcc@hc-sc.gc.ca).

Ottawa, June 9, 2016

Jurica Čapkun
Assistant Clerk of the Privy Council

Regulations Amending the Radiation Emitting Devices Regulations (Dental X-ray Equipment)

Amendments

1 Section 2 of Schedule I to the Radiation Emitting Devices Regulations (see footnote 1) is replaced by the following:

2 Dental X-ray equipment, being a radiation emitting device that is designed primarily for the examination of dental and maxillofacial structures in living humans and that has an X-ray generating tube designed to be used outside the mouth.

2 Section 12 of Schedule I to the Regulations is replaced by the following:

12 Diagnostic X-ray equipment, being a radiation emitting device that uses X-rays and is designed for the examination of the human body. It does not include dental X-ray equipment, photofluorographic X-ray equipment, radiation therapy simulators and computer-assisted tomographic equipment.

3 Part II of Schedule II to the Regulations is replaced by the following:

Part II

Dental X-Ray Equipment

Interpretation

Definitions

1 The following definitions apply in this Part.

aluminum means aluminum that has a degree of purity of 99.9% or higher and a density of 2.70 g/cm3. (aluminium)

aluminum equivalent means the attenuation equivalent of an object, expressed in thickness of aluminum. (équivalent en aluminium)

coefficient of variation means the ratio of the estimated standard deviation to the mean value of a series of measurements, determined by the formula

This image shows the coefficient of variation formula.

where

dental volumetric reconstruction means reconstruction of the three-dimensional attenuation distribution of all or part of the irradiated volume from a series of two-dimensional projections that are produced by an X-ray beam on an X-ray image receptor moving around the head of the patient. (reconstruction dentaire volumétrique)

deviation, in respect of any parameter of dental X-ray equipment, means the percentage error between the measured value and the value indicated either on the equipment or in the accompanying information. (variation)

digital X-ray image receptor means an X-ray image receptor whose conversion method is electrically powered. It includes both single-step direct conversion and multi-step indirect conversion. (récepteur numérique d'image radiologique)

effective image reception area means the part of the image reception area that is configured to receive an X-ray pattern that can be processed for display or storage. (surface receptrice de l'image efficace)

extra-oral, in relation to dental radiography, means that the X-ray image receptor is located outside the oral cavity. (extra-oral)

hand-held, in respect of transportable dental X-ray equipment, means that the equipment, once it is placed into service, is held in or supported by the hand. (portatif)

intra-oral, in relation to dental radiography, means that all or part of the X-ray image receptor is located inside the oral cavity. (intra-oral)

irradiation time means the duration of an irradiation, measured as the interval between the instant when the air kerma rate has risen for the first time to a value of 50% of the peak value and the instant when it has dropped for the last time below 50% of the peak value. (temps d'irradiation)

lead equivalent means the attenuation equivalent of an object, expressed in thickness of lead. (équivalent en plomb)

loading factor means a factor the value of which influences the X-ray tube load. It includes the following:

mode of operation means the technical state that is defined by a configuration of several predetermined loading factors and other settings for radiography or radioscopy, selectable simultaneously by the operation of a single control. (mode de fonctionnement)

rectification type means the process by which the X-ray generator of dental X-ray equipment converts high voltage to X-ray tube voltage. (type de redressement)

transportable, in relation to dental X-ray equipment, means that the equipment is constructed so that, once it is placed into service, it is capable of being moved from one place to another. (transportable)

X-ray field means the area on a surface that is intersected by a radiation beam whose boundary is determined by the points where the air kerma drops to 25% of the air kerma at the centre of the X-ray field. (champ de rayonnement X)

X-ray image receptor means a device that converts incident X-rays into a visible image or into a form that can be made into a visible image. (récepteur d'image radiologique)

Information and Labelling

Information

General requirements

2 The manufacturer, distributor and importer must ensure that all of the following information accompanies each piece of dental X-ray equipment:

Additional requirements — intra-oral equipment

3 The manufacturer, distributor and importer must ensure that all of the following additional information accompanies each piece of intra-oral dental X-ray equipment:

Additional requirements — extra-oral equipment

4 The manufacturer, distributor and importer must ensure that all of the following additional information accompanies each piece of extra-oral dental X-ray equipment:

Labelling

Presentation

5 The manufacturer, distributor and importer must ensure that all information that is required by this Part to be displayed on dental X-ray equipment is clearly legible, permanent and visible on the specified surface of the equipment when the equipment is fully assembled for use.

Function of controls

6 The manufacturer must ensure that all controls, warning lights and other indicators on the control panel are clearly labelled as to their function.

Information to be displayed

7 The manufacturer must ensure that all of the following information is displayed on dental X-ray equipment:

“WARNING: Hand-held operation increases operator radiation exposure due to proximity. See manufacturer safety information.”

« ATTENTION : Le fonctionnement en mode portatif augmente l'exposition de l'opérateur au rayonnement dû à la proximité. Consultez les renseignements de sécurité du fabricant. ».

Warning symbol

8 The X-ray warning symbol must have the following characteristics:

Construction and Functioning Standards

Control panel

9 The control panel of dental X-ray equipment must have all of the following features:

Irradiation switch

10 (1) Every irradiation switch of dental X-ray equipment must enable all of the following actions:

Hand-held intra-oral equipment

(2) Hand-held intra-oral equipment must also have a second irradiation switch that gives the operator the option to be at least 2 m from the X-ray source assembly when the equipment is supported on a stand.

Protected area

11 Dental X-ray equipment, except transportable equipment, must have a means to permit initiation of an irradiation by the operator from an area that is protected from radiation by structural shielding or distance.

Audible signal

12 Dental X-ray equipment must emit a signal, clearly audible to the operator, that indicates the end of an irradiation.

Line voltage fluctuations

13 Dental X-ray equipment must have a means appropriate to its rectification type to compensate for variations in X-ray tube voltage that are caused by line voltage fluctuations.

X-ray tube

14 The X-ray tube of dental X-ray equipment must be securely affixed to and aligned within the X-ray tube housing.

X-ray source assembly

15 The X-ray source assembly of dental X-ray equipment must maintain its required position or trajectory without drift or vibration during operation.

More than one X-ray source assembly

16 If more than one X-ray source assembly of dental X-ray equipment is controlled by the same control panel, the equipment must have a visual indicator on or near each X-ray tube housing that shows that the X-ray source assembly to which the indicator applies has been selected.

Filtration

17 (1) Dental X-ray equipment must have radiation-absorbing filters that meet the following requirements:

TABLE
Item

Column 1

X-ray Tube Voltage — Peak Value (kV)

Column 2

First Half-value Layer of Aluminum (mm)

1 60 2.2
2 70 2.5
3 80 2.9
4 90 3.2
5 100 3.6
6 110 3.9
7 120 4.3
8 130 4.7
9 140 5.0
10 150 5.4

Total filtration — intra-oral equipment

(2) Despite paragraph (1)(b), if intra-oral dental X-ray equipment operates at a maximum nominal X-ray tube voltage of 70 kV or less, it may instead have a total filtration of at least 1.5 mm aluminum equivalent.

Automatic exposure control

18 Dental X-ray equipment that has an automatic exposure control must have the following features:

Minimum distance — focal spot to skin

19 Dental X-ray equipment must have a device that limits the distance between the focal spot and the skin to at least the following distance:

Maximum deviation — focal spot to image receptor

20 If the distance between the focal spot and the X-ray image receptor of extra-oral dental X-ray equipment is adjustable, the equipment must have a visual indicator that shows the distance that is selected and that has a maximum deviation of 5%.

Beam-limiting device — extra-oral equipment

21 (1) Extra-oral dental X-ray equipment must have a beam-limiting device that meets the following requirements:

Beam-limiting device — intra-oral equipment

(2) Intra-oral dental X-ray equipment must have a beam-limiting device that meets the following requirements:

Incremental settings

22 (1) Subject to subsection (2), the range of settings for either the tube current, irradiation time or current time product of dental X-ray equipment must be in increments that are not greater than the decimal multiples and submultiples of 1.00, 1.25, 1.60, 2.00, 2.50, 3.20, 4.00, 5.00, 6.30 and 8.00.

Exception — intra-oral equipment

(2) In the case of intra-oral dental X-ray equipment, for irradiation times shorter than 0.08 s in one-peak high-voltage generators and two-peak high-voltage generators when it is not possible to provide all values that belong to the geometrical series within the range specified in subsection (1) because of the dependence on the pulsed nature of the supply mains, any missing values and the geometrical intervals that result from the absence of the missing values must be recognizable on the scale and must be explained in the instructions for use required by paragraph 2(e).

Modes of operation

23 Dental X-ray equipment that has more than one mode of operation must be constructed in accordance with the following requirements:

Current time product

24 For each selectable X-ray tube voltage of intra-oral dental X-ray equipment, the ratio of the maximum to minimum current time product settings must be at least 16:1.

Primary shielding

25 Extra-oral dental X-ray equipment must have primary protective shielding that completely overlaps the X-ray field and that provides the following minimum attenuation:

Attenuation equivalent

26 All parts of extra-oral dental X-ray equipment that are located in the path of the X-ray beam between the patient and the X-ray image receptor must, when taken together, have an attenuation of not more than 1.2 mm aluminum equivalent, as determined at the highest X-ray tube voltage of the equipment.

Hand-held intra-oral equipment

27 Hand-held intra-oral dental X-ray equipment must have all of the following features:

X-ray tube voltage

28 The minimum nominal X-ray tube voltage of dental X-ray equipment must be at least 60 kV.

Functioning requirement

29 Dental X-ray equipment must function in accordance with sections 30 to 35 under the following conditions:

Loading factors

30 (1) Subject to subsection (2), for any combination of loading factors, the measured value of a loading factor set out in column 1 of the table to this subsection must not deviate from the indicated value by more than the amount set out in column 2.

TABLE
Item

Column 1

Loading Factor

Column 2

Maximum Deviation

1 X-ray tube voltage 10%
2 X-ray tube current 20%
3 Irradiation time for extra-oral dental X-ray equipment 5% + 50 ms
4 Irradiation time for intra-oral dental X-ray equipment 5% or 20 ms, whichever is larger
5 Current time product for extra-oral dental X-ray equipment 10% + 0.2 mAs

Exception — intra-oral equipment

(2) In the case of intra-oral dental X-ray equipment that has a one-peak high-voltage generator, the maximum deviation set out in item 4, column 2, of the table to subsection (1), does not apply when the irradiation time is shorter than 0.1 s.

Coefficient of variation — air kerma

31 (1) For any combination of loading factors of dental X-ray equipment, the coefficient of variation of any five consecutive air kerma measurements — taken at the same point along the X-ray beam axis within a one-hour period — must be 0.05 or less.

Internally powered equipment

(2) Dental X-ray equipment that is internally powered must meet the requirement set out in subsection (1) over the whole range of usable charging levels of the internal supply.

Maximum deviation — air kerma

32 The maximum deviation of the air kerma must not exceed 50%.

Maximum deviation — dose area product

33 The maximum deviation of the dose area product for extra-oral dental X-ray equipment must not exceed 50%.

Air kerma — linearity

34 (1) Subject to subsection (2), for any selected value of X-ray tube voltage and over the whole range of current time product settings, for any two settings of the current time product that do not differ by more than a factor of 2, the quotients of the average of the air kerma measurements divided by the corresponding indicated current time product must not differ by more than 0.10 times their sum, as determined by the following formula, where X is the quotient of the average of the air kerma measurements divided by the indicated current time product, calculated at each of the two settings of current time product:

|X1 – X 2| ≤ 0.1(X1 + X2)

Exception — intra-oral equipment

(2) In the case of intra-oral dental X-ray equipment that has a one-peak high-voltage generator, the shortest irradiation time in the range of current time product settings is 80 ms.

Leakage radiation — loading state

35 (1) Leakage radiation from the X-ray source assembly of dental X-ray equipment must not exceed the following air kerma rate, when the equipment is operated at the nominal X-ray tube conditions of loading that correspond to the maximum specified energy input in one hour:

Detection area

(2) For the purpose of subsection (1), the air kerma rate must be averaged over a detection area of 100 cm2 — of which no linear dimension is greater than 20 cm — that is 1 m from the focal spot.

Leakage radiation — non-loading state — extra-oral equipment

(3) In the case of extra-oral dental X-ray equipment that is not in the loading state, the leakage radiation from the X-ray source assembly must not exceed an air kerma rate of 20 µGy/h.

Detection area

(4) For the purpose of subsection (3), the air kerma rate must be averaged over a detection area of 10 cm2 — of which no linear dimension is greater than 5 cm — that is 5 cm from any accessible surface of the equipment.

Transitional Provision

4 The Radiation Emitting Devices Regulations, as they read immediately before the coming into force of these Regulations, continue to apply in respect of dental X-ray equipment that was manufactured before the day on which these Regulations come into force.

Coming into Force

5 These Regulations come into force on the day that, in the sixth month after the month in which they are published in the Canada Gazette, Part II, has the same calendar number as the day on which they are published or, if that sixth month has no day with that number, the last day of that sixth month.